全部∨
搜索
科研活动
科技论文
当前位置: 首页>科研成果
论文名称
Energy response correction for an electronic personal dosimeter using the channel ratio method
发表刊物
Nuclear Science and Techniques
发表年度
2017
论文作者
Ping Huang、 You-De Zhang、Yong Yang、 Cong-Hua Zhang
承担单位
辐射所
收录情况
关键词
摘要
An electronic personal dosimeter mainly uses a Si-PIN photodiode as X- and gamma-ray detectors. The photon energy response of this instrument is inconsistent in the case of no correction, which seriously affects the accurate monitoring of personal dose equivalent H p (10)parameters for radiation workers. For this reason, in this paper we propose a method of combining composite screen detection technology, multichannel measurement technology, and the channel ratio method to achieve accurate measurement of the personal dose equivalent parameters. According to China National Standard GB/T 13161-2003 and National Verification Regulation JJG 1009-2006, the instrument was tested in the energy range between 48 keV and 1.25 MeV. The experimental results showed that the difference of energy response to 137 C s corrected by the new method was almost constant within ?6:0%, which fulfilled the ?30% requirement of GB/T 13161-2003 and JJG 1009-2006. Meanwhile, the method proposed obtained energy information regarding the radiation field.
业务咨询:
028-60828828
联系电话:
028-60828828
邮箱:
nimtt@nimtt.com
邮编:
610021
地址:
四川省成都市玉双路10号
Copyright©2013 中国测试技术研究院 版权所有 蜀ICP备 09003775号 网站总访问量(单位:次):506062 今日独立用户访问量(单位:个):5253 21561

川公网安备 51010802000911号

Copyright©2013 中国测试技术研究院 版权所有